How do I choose a thermocouple type?
Start with the instrument input and the continuous and peak temperatures. Then check the atmosphere, required tolerance and whether the sheath, junction and wiring can operate at their installed temperatures.
Compare thermocouple types by application
K, J, T, N and E are common base-metal types. R, S and B use platinum-based alloys. Start with the instrument input, normal and peak temperatures, and process atmosphere, then check the limits of the protected assembly.
| Type | Selection context |
|---|---|
| K | General industrial use, including ovens and process equipment, subject to the atmosphere and sheath. |
| J | Heated tooling and systems configured for Type J. Review protection for the iron leg in moisture or oxidizing conditions. |
| T | Often considered for measurements at lower temperatures. Verify the probe and lead construction for the environment. |
| N | Consider for base-metal service where stability at elevated temperatures is important. Check instrument support. |
| E | Produces a relatively large thermoelectric signal for a given temperature change. Confirm the intended measurement range and input. |
| R | Platinum–13% rhodium and platinum pair for furnace or kiln service. Confirm the Type R input, protection and insertion depth. |
| S | Platinum and platinum–10% rhodium pair for selected furnace and kiln measurements. Review protection tube material and mounting. |
| B | Two platinum–rhodium legs for selected high-temperature service. Check the measurement interval, tube support and terminal head temperature. |
Compare Type K and Type J
Type K uses nickel-based conductors and is widely used for general industrial measurement. Type J uses an iron positive conductor that may need protection in oxidizing or moisture-prone service. For a replacement, confirm which type the instrument is configured to read, then compare the full probe assembly limits.
Compare Type N and Type K
Both use nickel-based alloys. Type N is often considered when stability during extended exposure to elevated temperatures is important. Compare behavior for the actual thermal cycle and atmosphere, and configure the instrument and wiring for the selected type.
Compare Type R and Type S
Both use a platinum negative leg. The positive leg contains 13% rhodium in Type R and 10% in Type S. Each has its own voltage-to-temperature reference function. Confirm the required instrument input and the specified furnace protection tube, mount and wiring.
Read the ITS-90 reference intervals
These NIST reference functions describe thermoelectric voltage over a defined temperature interval. To specify a probe, confirm the instrument span and review the sheath or protection tube, atmosphere, exposure duration, insulation and lead connection at their installed temperatures.
| Type | NIST ITS-90 reference function interval |
|---|---|
| N | −270 to 1,300 °C |
| R | −50 to 1,768.1 °C |
| B | 0 to 1,820 °C |
Define the complete assembly
The alloy type is only one input. Sheath, insulation, junction, connector and extension wire can impose their own limits. Include expected peak temperature, probe depth, vibration and the mounting interface in a request.
Match the mounting to the sensing point
For a threaded port, specify the thread and measure the desired insertion from the installed reference face. For spring-loaded contact, describe the mating feature and installed travel. A drawing that identifies the sensing point helps connect these dimensions to the probe design.
Decide how to connect it
Match the extension cable and connector to the thermocouple type and instrument. Document any required terminal head, miniature plug or bare wire termination.